      <ArticleType>Pediatric Dentistry</ArticleType>
      <ArticleTitle>Comparative evaluation of the accuracy of Propex PixiTM and CanalProTM electronic apex locators in primary anterior teeth</ArticleTitle>
      <SubTitle/>
      <ArticleLanguage>English</ArticleLanguage>
      <ArticleOA>Y</ArticleOA>
      <FirstPage>44</FirstPage>
      <LastPage>50</LastPage>
      <AuthorList>
        <Author>
          <FirstName>Gaurav</FirstName>
          <LastName>Shukla</LastName>
          <AuthorLanguage>English</AuthorLanguage>
          <Affiliation/>
          <CorrespondingAuthor>N</CorrespondingAuthor>
          <ORCID/>
          <FirstName>Dinesh</FirstName>
          <LastName>Rao</LastName>
          <AuthorLanguage>English</AuthorLanguage>
          <Affiliation/>
          <CorrespondingAuthor>Y</CorrespondingAuthor>
          <ORCID/>
          <FirstName>Sunil</FirstName>
          <LastName>Panwar</LastName>
          <AuthorLanguage>English</AuthorLanguage>
          <Affiliation/>
          <CorrespondingAuthor>Y</CorrespondingAuthor>
          <ORCID/>
        </Author>
      </AuthorList>
      <DOI>10.54276/JUPD.2023.2107</DOI>
      <Abstract>Aim: To evaluate and compare in vitro, the accuracy of two different electronic apex locators, i.e., Propex PixiTM and CanalProTM in primary anterior teeth.&#13;
&#13;
Methodology: A total of 60 primary teeth sample has been selected using a convenience sampling method. Samples were divided into two experimental groups in which, Group 1 belong to Propex PixiTM apex locator, and Group 2 CanalProTM. The mean values with standard deviations of electronic working length by using Propex PixiTM and CanalProTM with that of stereomicroscopic actual working length were calculated and compared&#13;
&#13;
Results: There was statistically significant difference between values obtained by the two electronic apex locators and actual length Conclusion: A strong correlation was seen between the two electronic methods and AL and also in between the two EALs, showing the possibility of their use to measure the root canal length.</Abstract>
      <AbstractLanguage>English</AbstractLanguage>
      <Keywords>Working length, Electronic apex locators, Actual length, Propex PixiTM, CanalProTM</Keywords>
      <URLs>
        <Abstract>https://thejupd.com/ubijournal-v1copy/journals/abstract.php?article_id=14992&amp;title=Comparative evaluation of the accuracy of Propex PixiTM and CanalProTM electronic apex locators in primary anterior teeth</Abstract>